Giraffe PT

Single-shot Bragg peak measurement

Overview

Check your complete Bragg Peak depth dose curve distribution for single beam spot delivery

The Giraffe PT is a high-resolution multi-layer ionisation chamber, which is consisting of 180 large 12 cm diameter electrodes

  • Instant absolute range verification for weekly & monthly QA
  • 3x faster commissioning of your particle therapy rooms
  • Reconstruct the Bragg-peak for a single spot within seconds
  • Verify the clinical range of up to 230 MeV

Instant daily PBS verification within one second

  • Check your complete Bragg Peak depth dose curve distribution for single beam spot delivery

Minimise loss of data collection

  • Safe time with reduced effort to correct data at TPS level. Verify the TPS library periodically without needing to set up the water tank.

Technical specs

  • Water equivalent material for equivalent scattering properties
  • Measure ranges with an accuracy of 0.5 mm compared to measurements in water (Bortfeld fit)
  • Verify the full clinical range with proton beam energies up to 230 MeV
  • 180 chambers with 2 mm pitch
  • 439 (L) x 17 (W) x 19.5 (H)