How Dosimetry Check works


To compute the dose to the patient from a treatment portal, pencils are first created that cover the treatment field and patient. The area of a beam field is simply divided up into small pixels (in a plane perpendicular to the beam), each pixel being a separate pencil. Whether a pencil beam algorithm or a full superposition algorithm, it is necessary to trace each pencil through the patient as illustrated in the below image where we have lighted up every other pencil.

fanlines

Typically the density of the material the beam is going through is noted and one computes terma along each ray. However, to compute the dose to the patient one must know the intensity of radiation that reaches each ray.

In a planning system to know the intensity that is reaching each pencil on the surface of the patient, a source model is used that can compute the intensity down stream from the collimation system on any point on a plane perpendicular to the central ray, as illustrated in the below image. The source model must take into account everything, the flattening filter, the collimator system, the multi-leaf system, wedges, and any dynamic movements of those things to arrive at the final intensity that reaches each pencil.

beamgrid

Measured source model

Dosimetry Check does not have a source model, which is the whole point of Dosimetry Check. In Dosimetry Check, the source model is replaced with a measured source model. If you were to put a piece of x-ray film on the couch and expose to the treatment portal straight down for the whole beam on time, you will have a picture of the field intensity that takes into account everything, such as leaf leakage, since you have just measured it all. In Dosimetry Check you can use the EPID or a diode or ion chamber array for greater convenience. The distance to the patient surface for each pencil is known from the CT scans and beam position.


The beam film (taken BEFORE or without the patient) gets darker the more radiation a spot gets. As noted above, the port image without the patient in the beam is a record of how much radiation each pixel gets. It is necessary to convert this information to a unit that will make it possible to compute the dose in centiGray to the patient from each pencil. Each spot on the image is mapped to the monitor units that would produce the same darkness at the center of a 10x10 cm field size (i.e. the field size that you calibrate the accelerator to). We call this resultant number the "relative monitor units" (RMU).

exp_epid_fl

Shown left is an EPID image of a modulated field for a head and neck case after mapping to RMU units (where white is more radiation).


vfilmcal

To accomplish this mapping you must run a calibration curve of the pixel value at the center of a 10x10 cm field versus monitor units. This curve is used to map the pivel values on the beam images above to RMU. However, for an EPID, ion chamber or diode array, integration is a linear process with zero intercept. So one can use a single measurement of a 10x10 cm field to determine the same calibration line. This means only one 10x10 field need be measured with a clinical case to map the fluence images to RMU values. But an EPID or similar device may generate internal scatter within the device. This is corrected for by doing a deconvolution with the inverse of the point spread function of the device to arrive at in air fluence in RMU.


epid_cal

Once converted to RMU, Dosimetry Check can compute the dose from a beam to the patient in centiGray. Here, zero RMU is black, whiter is larger RMU. The pixels on the RMU image are used as "weights" for the corresponding pencil beams. The fluence map that we have derived here from a calibrated image of the beam completely determines the dose to the patient. This process is similiar to how a planning system computes the dose, except here we are using the measured fluence instead of computing the same from knowledge and models of what kind of things modulate the beam. By starting with measurement instead of a model, we verify the dose and dose distribution that the patient receives.

brmu5

The dose computed by Dosimetry Check may be compared directly to that computed by the planning system. ADAC IMRT plan (green) versus Dosimetry Check (magenta) with inhomogeneity corrections on, courtsey of Dr. Tianyou Xue.